GTX-Plus Series User’s Manual REV 2.1

21

SELECT RECORD

SAVED RECORD: 002/200
MAX RECORD COUNT: 20

BEGIN RECORD: 001

END RECORD: 002

Figure 15c Select record Menu

Press UP and DOWN key to select the record, press and hold to

scroll select the record. Press LEFT and RIGHT key to change cursor

between BEGIN RECORD and END RECORD, press and hold to scroll
change.

*BEGIN RECORD and END RECORD must be not equal 0.

*BEGIN RECORD must be less than END RECORD.

*The maximum of record for calculation must be not over 20.

*Measurement unit and direction of each record for calculation

must be same.

Press ENTER key to calculate the statistic values, The display will

show calculation result page. Press ESC key to return to Database menu
page.



Figure 15d Statistics result Menu

*Pass and Fail count calculation are depend on last setting of

UPPER LEVEL and LOWER LEVEL.

3) CALIBRATION This is used by service technicians when calibrating
the gauge. Contract your Nextech distributor for details.

4) DATE&TIME This use to set date and time. To set date and time, Go

to main menu page and press UP and DOWN key to move cursor point to
DATE&TIME and press ENTER key the display will show Date&Time menu

page. Press ESC key to return to main menu page.

MAX: 1.0860 kgf
MIN: 0.7360 kgf
MEAN: 0.9110 kgf
STD DEV: 0.2474 kgf
COV: 27.166 %

PASS: 02 ( 100.00 %)
FAIL: 00 (0.0000 %)