System problems

The electrocardiograph won't turn on when it is plugged in

Causes

The AC power connection is faulty.

An AC fuse is blown.

There is no AC power.

Actions

Check the AC power source.

Check the AC fuses.

The electrocardiograph won't turn on when it is unplugged

Causes

The battery is disconnected or incorrectly connected.

The battery is low, not charging, depleted, or bad.

Actions

Check the battery connections.

Recharge the battery.

Replace the battery.

The electrocardiograph shuts down during printing

Causes

The battery is low or bad.

Actions

Recharge the battery.

Replace the battery.

The electrocardiograph prints fewer than 10 reports on a full battery

charge

Causes

The battery is degraded.

Actions

Replace the battery.