Maintenance
Cleaning the equipment
WARNING Keep the electrocardiograph, reusable electrodes, and the
patient cable clean. Patient contact with contaminated equipment can
spread infection.
CAUTION Never let soap or water come into contact with the
electrocardiograph’s internal printer, connectors, or jacks.
CAUTION Never immerse the electrocardiograph or the patient cable in
liquid. Never autoclave or steam clean the electrocardiograph or the patient
cable. Never pour alcohol directly on the electrocardiograph or the patient
cable, and never soak any components in alcohol. If any liquid enters the
electrocardiograph, remove the electrocardiograph from service, and have
it inspected by a qualified service person before using it again.
Note
The patient cables should be cleaned after each use.
Clean on a routine basis according to your facility's protocols and standards or local
regulations.
The following agents are compatible with the electrocardiograph:
•
70 percent isopropyl alcohol
•
10 percent chlorine bleach solution
CAUTION When cleaning the device, avoid using cloths or solutions that
include quaternary ammonium compounds (ammonium chlorides) or
glutaraldehyde-based disinfectants.
Note
Disinfect according to your facility's protocols and standards or local
regulations.
To clean the equipment (monthly or more often if needed)
1.
Disconnect the power plug from the AC outlet.