2 Contour analysis

2.1 Contour parameters

All parameters that are used in the diagnostic criteria of the contour classification are measured in the
representative P-QRS-T complex. The lead-independent, overall parameters are presented in Table 1.

Table 1. Lead-independent parameters for the contour classification.

Name

Description

Heart rate

Ventricular rate (in beats per minute, bpm)

P axis

Axis of the P wave (in degrees, from

–180 to 180)

P duration

Duration of the P wave (in ms)

PR interval

Duration of the PR interval (in ms)

QRS axis

Axis of the QRS complex (in degrees, from

180 to 180)

QRS duration

Duration of the QRS complex (in ms).

Corrected QT interval

QT interval corrected for heart rate according to Bazett’s formula:
QTc = QT *

(HR/60) (in ms)