3.16 Group 13: Irregular rhythms with P/QRS > 1.2 and PR range > 30 ms

Say:

“supraventricular (sinus?) tachycardia with second degree AV block at”

“ variable conduction ratio”

if:

270

shortest PP interval < LLN


Say:

“sinus rhythm with second degree AV block at variable conduction ratio”

if:

shortest PP interval

LLN

3.17 Group 14: Irregular rhythms with P/QRS > 1.0 and PR range

30 ms

The irregularity of the ventricular rhythm combined with the constancy of the PR interval and a higher
number of P waves than QRS complexes implies that either a second degree AV block with constant
preceding conduction times (Mobitz type II) or advanced AV block with varying conduction ratios is
present. In case of Mobitz type II AV block, the longer RR intervals will most likely be sandwiched
between shorter RR intervals. This is unlikely to occur in cases with advanced AV block with varying
conduction ratios.

Say:

“atrial flutter”

if:

shortest PP interval < 270 ms


Say:

“atrial tachycardia”

if:

270

shortest PP interval < 370 ms


Say:

“sinus tachycardia”

if:

370

shortest PP interval < LLN


Say:

“sinus rhythm”

if:

shortest PP interval

LLN


Say:

“with second degree AV block, type II (Mobitz II)”

if:

varying RR intervals fulfilling Mobitz II criteria


Say:

“with second degree AV block at variable conduction ratio”

if:

RR intervals not fulfilling Mobitz II criteria