beautypg.com

Cardioperfect workstation – Welch Allyn Cardioperfect Workstation, PEDMEANS ECG Interpretation Module - Physicians - User Manual User Manual

Page 39

background image

CardioPerfect Workstation

PEDMEANS ECG Interpretation Module Physicians Manual

DIR 80015051 Rev. C

39 / 55

3.13 Group 10: Irregular rhythms with 0.15 < P/QRS

0.9 and PR range

30

ms

First, a test is made on the alternation of RR intervals with and without a P wave to rule out a
bigeminal rhythm due to sinus rhythm with atrial or AV junctional premature complexes. If a bigeminal
rhythm is found the program halts. If not, tests are performed on each RR interval in which no P wave
has been detected. If the interval is shortened, a premature supraventricular complex is assumed; if it
is prolonged, an AV junctional escape is assumed. The shortened or prolonged RR intervals are
deleted after they have been analyzed and the program continues with the logic for rhythms with
constant PR intervals.

Say:

“bigeminy: sinus rhythm with alternate premature supraventricular complexes”

if:

P/QRS ratio

0.4

and

test bigeminy passed


Say:

“premature supraventricular complexes”

if:

test PSVC passed

and

test bigeminy failed


Say:

“AV junctional escapes, cause? eg SA block or AV block?”

if:

test escape passed

and

test bigeminy failed


Say:

“consider premature supraventricular complexes”

if:

test bigeminy failed

and

test PSVC failed

and

test escape failed