3.10 Group 7: Regular rhythms with P/QRS > 1.0 and PR range > 30 ms
Say:
“atrial tachycardia”
if:
270
shortest PP interval < 370 ms
Say:
“sinus tachycardia”
if:
370
shortest PP interval < LLN
Say:
“sinus rhythm”
if:
shortest PP interval
LLN
Say:
“with second degree AV block at 2:1 conduction ratio”
if:
shortest PR interval > 120 ms
and
1.75 < P/QRS ratio
2.0
If the criteria for a 2:1 AV block are not fulfilled, an AV dissociation due to block or interference is
assumed.
Say:
“with complete AV block”
“idioventricular escape rhythm”
if:
QRS duration
ULN + 20 ms
and
heart rate < LLN
– 20 BPM
Say:
“with complete AV block”
“accelerated idioventricular escape rhythm, consider”
“AV junctional rhythm with aberrant ventricular conduction”
if:
QRS duration
ULN + 20 ms
and
LLN
– 20 BPM
heart rate < LLN
Say:
“with block and/or interference in the AV junction”
“accelerated AV junctional rhythm with aberrant ventricular conduction,”
“consider ventricular tachycardia”
if:
QRS duration
ULN + 20 ms
and
LLN
heart rate < ULN
Say:
“with block and/or interference in the AV junction”
“AV junctional tachycardia with aberrant ventricular conduction,”
“consider ventricular tachycardia”
if:
QRS duration
ULN + 20 ms
and
heart rate
ULN
Say:
“with complete AV block”
“AV junctional escape rhythm”
if:
QRS duration < ULN + 20 ms
and
heart rate < LLN
Say:
“with block and/or interference in the AV junction”
“accelerated AV junctional rhythm”
if:
QRS duration < ULN + 20 ms
and
LLN
heart rate < ULN
Say:
“with block and/or interference in the AV junction”
“AV junctional tachycardia”
if:
QRS duration < ULN + 20 ms
and
heart rate
ULN