beautypg.com

Cardioperfect workstation – Welch Allyn Cardioperfect Workstation, PEDMEANS ECG Interpretation Module - Physicians - User Manual User Manual

Page 37

background image

CardioPerfect Workstation

PEDMEANS ECG Interpretation Module Physicians Manual

DIR 80015051 Rev. C

37 / 55

3.10 Group 7: Regular rhythms with P/QRS > 1.0 and PR range > 30 ms

Say:

“atrial tachycardia”

if:

270

shortest PP interval < 370 ms


Say:

“sinus tachycardia”

if:

370

shortest PP interval < LLN


Say:

“sinus rhythm”

if:

shortest PP interval

LLN


Say:

“with second degree AV block at 2:1 conduction ratio”

if:

shortest PR interval > 120 ms

and

1.75 < P/QRS ratio

2.0


If the criteria for a 2:1 AV block are not fulfilled, an AV dissociation due to block or interference is
assumed.

Say:

“with complete AV block”

“idioventricular escape rhythm”

if:

QRS duration

ULN + 20 ms

and

heart rate < LLN

– 20 BPM


Say:

“with complete AV block”

“accelerated idioventricular escape rhythm, consider”

“AV junctional rhythm with aberrant ventricular conduction”

if:

QRS duration

ULN + 20 ms

and

LLN

– 20 BPM

heart rate < LLN


Say:

“with block and/or interference in the AV junction”

“accelerated AV junctional rhythm with aberrant ventricular conduction,”

“consider ventricular tachycardia”

if:

QRS duration

ULN + 20 ms

and

LLN

heart rate < ULN


Say:

“with block and/or interference in the AV junction”

“AV junctional tachycardia with aberrant ventricular conduction,”

“consider ventricular tachycardia”

if:

QRS duration

ULN + 20 ms

and

heart rate

ULN


Say:

“with complete AV block”

“AV junctional escape rhythm”

if:

QRS duration < ULN + 20 ms

and

heart rate < LLN


Say:

“with block and/or interference in the AV junction”

“accelerated AV junctional rhythm”

if:

QRS duration < ULN + 20 ms

and

LLN

heart rate < ULN


Say:

“with block and/or interference in the AV junction”

“AV junctional tachycardia”

if:

QRS duration < ULN + 20 ms

and

heart rate

ULN