beautypg.com

Cardioperfect workstation, 17 combination of statements – Welch Allyn Cardioperfect Workstation, PEDMEANS ECG Interpretation Module - Physicians - User Manual User Manual

Page 24

background image

CardioPerfect Workstation

PEDMEANS ECG Interpretation Module Physicians Manual

DIR 80015051 Rev. C

24 / 55

2.17 Combination of statements

If repolarization disturbances at different localizations have been detected, the program tries to
combine separate localizations in one statement. When “right precordial”, “mid precordial”, and “left
precordial” repolarization disturbances are present together, the statement “extensive precordial
repolarization disturbance” is made.

If infarction is present at different locations, the program attempts to generate a combined infarction
statement. The following combinations can be made:
“inferior”+“lateral”

“inferolateral”

“anterior”+“lateral”

“anterolateral”

“anterior”+“septal”

“anteroseptal”

“anterior”+“septal”+“lateral”

“extensive anterior”

“anterior”+“septal”+“lateral”+“high lateral”

“extensive anterior and highlateral”