Single use sigmoidoscopes and anoscopes, Sterilization – Welch Allyn Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, Accessories - Cleaning, Disinfection, and Sterilization - User Manual User Manual
Page 9

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories
ENGLISH
7
Sterilization
1.
Follow steps 1 through 10 of the Cleaning instructions.
2.
Sterilize using a STERRAD
®
100S for one 55 minute cycle. Validate any other
sterilization equipment and load configurations.
Single Use Sigmoidoscopes and Anoscopes
Sterilization
1.
Remove the device from the plastic bag before sterilizing.
2.
Sterilize using a STERRAD 100S for one 55 minute cycle. Validate any other
sterilization equipment.
Single Use Anoscope
Part Numbers:
Single Use Sigmoidoscope
Part Numbers:
53110
53130
53110-B
53130-B
53110-JP
53130-F
53130-JP
Caution
Welch Allyn recommends the use of the Single Use
Insufflation Bulb Filter (
30210) for all applicable endoscopic
procedures. The filter does not eliminate the need to properly process
the insufflation bulb with each use, but restricts internal contamination
and aspiration of fluids.
Caution Sterilization process is not valid for 53130-L and 53130-LF.
These single use devices are lubricated and cannot be sterilized.
Caution These devices are not reusable, but can be sterilized one time
prior to single use.
53110
53130