Cleaning, Disinfection, and Sterilization Instructions
Rigid Sigmoidoscopes, Anoscopes, and Light Carrier
*Suction Tubes not illustrated
Cleaning
1.
Prepare an enzymatic detergent formulated for endoscopic instruments according to
the manufacturer’s instructions.These instructions were validated using ENZOL
®
/
CIDEZYME
®
.
2.
Submerge the item in the cleaning solution.
3.
Using a soft bristle brush, scrub the item submerged in the cleaning solution for a
minimum of 5 minutes then rinse with sterile water for 5 minutes.
High Level Disinfection
1.
Follow the Cleaning instructions above.
2.
Using CIDEX
®
OPA (0.55% OPA, 14-day maximum reuse), soak the item for a
minimum of 12 minutes.
3.
Perform 3 sequential rinses of the item for a minimum of 5 minutes each with sterile
water.
Anoscope
Part Numbers:
Sigmoidoscope
Part Numbers:
Light Carrier
Part Numbers:
37019
38614
32010
32820
36019
37023
38619
32020
32830
38700
37027
38622
32410
33220
Suction Tube*
Part Numbers:
38108
38900
32420
33620
38114
39614
32810
33830
30130
38119
39619
30140
38122
39622
Note
For all steps, remove obturator for separate cleaning and open viewing window.
Caution Welch Allyn recommends the use of the Single Use Insufflation Bulb
Filter (
30210) for all applicable endoscopic procedures. The filter does not
eliminate the need to properly process the insufflation bulb with each use, but
restricts internal contamination and aspiration of fluids.