Cleaning, Disinfection, and Sterilization Instructions

Rigid Sigmoidoscopes, Anoscopes, and Light Carrier

*Suction Tubes not illustrated

Cleaning

1.

Prepare an enzymatic detergent formulated for endoscopic instruments according to
the manufacturer’s instructions.These instructions were validated using ENZOL

®

/

CIDEZYME

®

.

2.

Submerge the item in the cleaning solution.

3.

Using a soft bristle brush, scrub the item submerged in the cleaning solution for a
minimum of 5 minutes then rinse with sterile water for 5 minutes.

High Level Disinfection

1.

Follow the Cleaning instructions above.

2.

Using CIDEX

®

OPA (0.55% OPA, 14-day maximum reuse), soak the item for a

minimum of 12 minutes.

3.

Perform 3 sequential rinses of the item for a minimum of 5 minutes each with sterile
water.

Anoscope

Part Numbers:

Sigmoidoscope

Part Numbers:

Light Carrier

Part Numbers:

37019

38614

32010

32820

36019

37023

38619

32020

32830

38700

37027

38622

32410

33220

Suction Tube*
Part Numbers:

38108

38900

32420

33620

38114

39614

32810

33830

30130

38119

39619

30140

38122

39622

Note

For all steps, remove obturator for separate cleaning and open viewing window.

Caution Welch Allyn recommends the use of the Single Use Insufflation Bulb
Filter (

30210) for all applicable endoscopic procedures. The filter does not

eliminate the need to properly process the insufflation bulb with each use, but
restricts internal contamination and aspiration of fluids.

37023

32820

36019