Cleaning, Disinfection, and Sterilization Instructions

4.

After performing a reprocessing procedure, wipe the lamp with a soft, dry, lint free
cloth.

5.

Re-attach the clean sleeve to the handle by screwing in a clockwise direction.

6.

Re-connect the power supply cord to the light handle and verify that the unit works by
plugging it in. If the lamp fails to illuminate follow the troubleshooting instructions
located in light handle manual (

700791) supplied with your handle.

Cleaning (73210, 73211 and 73500)

1.

Prepare an enzymatic detergent formulated for endoscopic instruments according to
the manufacturer’s instructions.These instructions were validated using ENZOL/
CIDEZYME.

2.

Submerge the rectal light handle outer sleeve in the cleaning solution.

3.

Using a soft bristle brush, scrub the item submerged in the cleaning solution for a
minimum of 5 minutes then rinse with sterile water for 5 minutes.

4.

Wet the rectal light handle body and cord with the cleaning solution.

5.

Using a soft bristle brush, scrub the cord with the cleaning solution for a minimum of
5 minutes then rinse the cord with sterile water for 5 minutes.

Intermediate Level Disinfection (73210, 73211 and 73500)

1.

Follow the Cleaning instructions above.

2.

Using CaviWipes

, wipe all item surfaces and keep wet with CaviWipe solution for a

minimum of 5 minutes.

3.

Allow the item surface to dry for a minimum of 10 minutes.

Sterilization (73211 ONLY)

1.

Follow the Cleaning instructions above.

2.

Perform one autoclave cycle that includes 6 minutes at 270° F (132° C) and a cool
down period of at least 15 minutes. These instructions were validated using a gravity
autoclave. Validate any other sterilization equipment and load configurations.

Caution Autoclaving the light handle may reduce the product’s functional life.