Power supply, Cleaning the power cord, Intermediate level disinfection of the power cord – Welch Allyn Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, Accessories - Cleaning, Disinfection, and Sterilization - User Manual User Manual
Page 10
8
ENGLISH
Cleaning, Disinfection, and Sterilization Instructions
Power Supply
The following cleaning instructions apply to the Power Cord ONLY:
1.
Disconnect the power supply cord from the light handle.
2.
Disconnect the power supply from its wall receptacle.
Cleaning the Power Cord
1.
Prepare an enzymatic detergent formulated for endoscopic instruments according to
the manufacturer’s instructions. These instructions were validated using ENZOL/
CIDEZYME.
2.
Wet the cord with the cleaning solution.
3.
Using a soft bristle brush, scrub the cord with the cleaning solution for a minimum of
5 minutes then rinse the cord with sterile water for 5 minutes.
Intermediate Level Disinfection of the Power Cord
1.
Follow the Cleaning instructions above.
2.
Using CaviWipes, wipe the cord and keep wet with CaviWipe solution for a minimum
of 5 minutes.
3.
Allow the cord surface to dry for a minimum of 10 minutes.
Caution
Do not immerse the power supply in any solution or sterilize. Doing so
may damage the power supply and void the warranty.
Caution The power supply can only be wiped with a soft, dry, lint free cloth or
scrubbed with a dry brush.
Part Numbers:
73305, 73322, 73324, 73326