Rechargeable Handle

Cleaning

1.

Using CaviWipes, wipe the surface to remove all gross debris.

2.

Using a soft bristle brush, scrub the handle for a minimum of 5 minutes then repeat
wipe with a second CaviWipe.

3.

Allow the handle surface to dry for a minimum of 10 minutes.

Intermediate Level Disinfection

1.

Follow the Cleaning instructions above.

2.

Using CaviWipes, wipe the handle and keep wet with CaviWipe solution for a
minimum of 5 minutes.

3.

Allow the handle surface to dry for a minimum of 10 minutes.

Rheostat
(Brightness
Control)

Part Number 71000-A

Caution

Do not saturate the rheostat area. Excess solution entering the

assembly may damageinternal components in the handle.