Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories

ENGLISH

3

4.

Allow the item to dry for 24 hours or until no moisture remains in the device. Any
moisture in the device could occlude the insufflation bulb filter.

Sterilization

1.

Follow the Cleaning instructions above.

2.

Perform one autoclave cycle that includes 6 minutes at 270° F (132° C) and a cool
down period of at least 15 minutes. These instructions were validated using a gravity
autoclave. Validate any other sterilization equipment and load configurations.

Light Handle Assembly and Handle Adapter

Assembly and Disassembly Instructions for 73210 and 73211

1.

Disconnect the light handle from the power supply cord and the device.

2.

Unscrew the outer sleeve of the handle by turning it counterclockwise.

3.

Follow one of the reprocessing methods described below.

Part Numbers 73210 and 73211

(including 07800-U lamp)

Part Number 73500

(including 03100-U lamp)

WARNING Allow the lamp to cool for at least 5 minutes before handling.

Note

Leave the lamp installed while reprocessing 73210, 73211 and 73500.

Caution Do not immerse the rectal light handle body and cord in any solution.
Doing so may damage the rectal light handle body and cord and void the
warranty.

Rectal Light Handle Body
and Cord

CCW

CW

Rectal Light
Handle Outer
Sleeve