Instruction Manual - TORBAL AGC Series
(B) Parts Counting
1. Place an empty container on the pan and press the “T” key to tare.
2. After taring the container, the scale will display the sample size setting which was configured in
the Mode Menu. For example SPL-10 will be displayed, indicating that the current sample size
setting is 10 pieces. To accept the sample size press the “T” (YES) key. To change the sample
size use the 0 (NO) key.
To Accept
To Change
The sample size selection can range from 1 to 100 pieces. By applying short single presses to the
0 (NO) key the sample size will change in increments of one. By holding the 0 (NO) key
pressed, the sample size will change in increments of 5. To select a sample size of 99, press and hold the
0 key until the display reads 95, and apply four single presses until the sample size reads 99. Once
the desired sample size is displayed press the “T” (YES) key to accept and confirm the selection.