Valor™ 7000, Series en-13, 2 cal menu – Rice Lake Valor 7000, Ohaus - Instruction Manual User Manual
Page 15

Valor™ 7000
Series EN-13
4.2 Cal Menu
Enter this menu to perform calibrations.
Initial Calibration
When the scale is operated for the first time, a span calibration is recommended to ensure
accurate weighing results. Before performing the calibration, be sure to have the appropriate
calibration weights as listed in table 4-1. Ensure that the LFT switch / calibration lock is set to
unlock position. See figure 5-1.
Or adjust the GEO setting according to your location (see table 4-2).
Procedure:
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released, the display
will show [C.A.L]. Press Yes to accept. [SpaN] will then be shown. Press Yes to begin the span
calibration. [ 0
kg
] will be displayed. Press Yes to accept. [--C--] will be displayed while zero
reading is stored. Next, the display shows the calibration weight value. Place the specified
calibration mass on the pan. Press Yes to accept the weight or No to select an alternate weight.
[--C--] will be displayed while the reading is stored. The display will show [done] if the calibration
was successful. The scale returns to the previous application mode and is ready for use.
TABLE 4-1
Required Span Calibration Mass (sold separately)
Capacity Mass
*
Capacity Mass
*
1500g 1.5kg / 3lb
15000g 15kg / 30lb
3000g 3kg / 6lb
30000g 30kg / 60lb
6000g 6kg / 15lb
Notes: When active unit is g or kg, the calibrating unit will be in kg.
When active unit is lb, oz or lb:oz, the calibrating unit will be in lb.
For linearity calibration, the calibration Mass is fixed.
Mid-point is always half of full capacity.
Span
[SpaN]
Initiates a span calibration procedure (zero and span)
Lin
[LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
GEO
[GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Settings from 0 to 31 are available with 12 being the default.
Refer to table 4-2 to determine the GEO factor that corresponds to your location.
End Cal
[End]
Advance to the next menu or return to the top of the current menu.
Span:
Perform
Linearity:
Perform
Geographic:
Adjustment:
Set 0...12...31*
End Calibration: Exit menu
*Bold always represents factory default value.