0 service information, 1 troubleshooting guide, Table 4-1. troubleshooting table – Rice Lake Summit 3000 Low-Profile Package User Manual
Page 14: 2 periodic maintenance, 3 load cell replacement, Table 4-2. load cell replacement numbers

10
Summit 3000 Installation Manual
4.0
Service Information
The following sections describe basic service and maintenance procedures.
4.1
Troubleshooting Guide
The following table lists some of the common problems and their suggested solutions.
4.2
Periodic Maintenance
The space between the platform side and pit frame, and the surface beneath the platform must be periodically
cleaned to prevent debris build up. More frequent cleaning of these areas is necessary with scales mounted in
pits.
Do not attempt to use scales with load cells that are not hermetically sealed in washdown applications. Water
damage is a common cause of failure in non-hermetically-sealed load cells. Use care with high pressure steam
washdowns for hermetically-sealed load cells. The steam may not damage the load cells, but the elevated
temperatures may cause incorrect readings until the unit cools to room temperature.
4.3
Load Cell Replacement
Replacement load cells can be ordered from Rice Lake Weighing Systems according to the part numbers listed
below.
Problem
Symptom Description
Solution
System does not
operate - no display
Power disconnected
Check and reconnect
Indicator fuse blown
Replace the fuse. Check for cause
Interface cable cut or disconnected
Repair
Signal leads incorrectly installed at the
indicator
Install according to indicator installation
manual
Display stays at zero
Indicator faulty
Service indicator
Load cell connections faulty
Check cable connections in junction box and
at indicator
Erratic weight
readings
Vibration near scale
Remove source of vibration or move scale
Platform not level to within 1/4 in.
Level scale by adjusting feet or shimming if
necessary
Load cell or cable water damage
Replace
Debris under load cells or platform
Clean
Indicator faulty
Use simulator to test indicator for stability.
Service indicator
Consistently high or
low weights
Indicator not properly adjusted to zero
Zero the indicator according to the indicator
manual
Platform binding
Obtain adequate clearance for free platform
movement
Indicator not calibrated
Calibrate according to indicator manual and
Section 3.3 on page 8
Load cells faulty
Test and replace load cells if necessary
Feet touching deck underside
Adjust feet downward to provide clearance
Table 4-1. Troubleshooting Table
Rice Lake PN
Summit Model #
Load Rating
30106
2K models
1000 lb (453.6 kg)
Table 4-2. Load Cell Replacement Numbers