32
RoughDeck HE AutoLift
for repeatability. If necessary, repeat steps 4 and 5.
7. Pull excess cable out of the enclosure and tighten the cord grip assemblies with a wrench. To be watertight,
each cord grip must be tightened so the rubber sleeve begins to protrude from the hub.
8. Unused hubs must be properly plugged to prevent moisture entry. See the Electronic Replacement Parts
catalog to order extra hole plugs.
9. Remove the desiccant from the plastic bag, and insert the desiccant bag into the junction box before
closing. Inspect the desiccant during normal service and change the desiccant as needed.
10. Replace the cover and torque the cover screws in an alternating pattern 15 in/lbs to be certain the gasket is
compressed equally in all locations.
LOAD CELL CABLE
STEEL COVER
FLOOR
SCREW
LOAD CELL CABLE
STEEL COVER
When the interface cable is protected and in its final position, complete connections to the indicator. See indicator
installation manual for wiring information.
If necessary, trim corners as described in Section 3.2 on page 6. Check all strain relief fittings for tightness. Slide
the junction box assembly into the cutout and secure it with the two #10 x
3
/
8
" screws provided.
2.6
Access Ramps
Access ramps for the RoughDeck AX are designed to bolt to the floor, with built-in mounting plates that attach to
the scale feet. When used with access ramps, side movement of the scale is automatically eliminated, and no other
mounting plates are necessary.
Access ramps can only be attached to the scale on one of the two scale sides that are perpendicular to the
longitudinal axis of the load cells. For example, the scale shown in Figure 2-4 could have an access ramp on the
left side as shown, and/or on the right side. The top and bottom sides, however, will not accept the ramp mounting
plates.
Longitudinal axis
of load cells
Figure 2-4. Access Ramps