0 maintenance, 1 calibration, Maintenance – Rice Lake Farm Bars - KLU Series Universal Weighbars User Manual
Page 12

8
KLU Series Universal Weighbars Operator’s Manual
5.0
Maintenance
The most common type of problem that can occur is the load is not supported completely by the load bar pads. On
a regular basis check around and under the weigh apparatus to see if any debris has collected near the scale. Any
ice, dirt, mud or manure that builds up around the weigh apparatus can cause inaccurate readings. The scale should
be kept clean to ensure proper operation.
It should also be noted that an excess of debris on top of the scale could also cause problems. Depending on the
type of scale you have, there may be a limited range of weight you can “zero” off the scale. Always keep the build-
up of material on top of the scale to a minimum.
Lubrication of the bearings in your scale is also important. The bearings are located in the bearing collar, as shown
in the parts list diagrams in Figure 4-1 and Figure 4-2. Inject grease into the bearings using a syringe at least once
every two years. Use good high-pressure grease to lubricate the bearings completely. Do not use an excess of
grease – it should lubricate the bearing without bleeding around the foam seal.
5.1
Calibration
All Load cells are calibrated at the factory to reduce installation errors. Over the course of time, however, the load
cells may drift out of calibration slightly. Normally when a scale's load cells go out of calibration, the digital
indicator is adjusted to compensate for this.
The KLU Series Loadbars can be calibrated using the front panel, EDP commands, or the Revolution
®
configuration utility. Each method consists of the following steps:
•
Zero calibration
•
Entering the test weight value
•
Span calibration
•
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
XXXXXXX
XXXXXXX
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
Display and edit
test weight value
WVAL
*CAL*
Display and edit
span calibration
A/D count value
WSPAN
VERS
PROGRM
PFORMT
SERIAL
CALIBR
CONFIG
FORMAT
Press Enter to
remove offset from
zero and span
calibrations
REZERO
*CAL*
Figure 5-1. Calibration (CALIBR) Menu
Front Panel Calibration
To calibrate the indicator using the front panel, do the following:
1. Place the indicator in setup mode (display reads
CONFIG
) and remove all weight from the scale platform.
If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.
2. Press until the display reads
CALIBR
(see Figure 5-1). Press to go to zero calibration (
WZERO
).
3. With
WZERO
displayed, press
to calibrate zero. The indicator displays
*CAL*
while calibration is in
progress. When complete, the A/D count for the zero calibration is displayed. Press
again to save the
zero calibration value and go to the next prompt (
WVAL
).
4. With
WVAL
displayed, place test weights on the scale and press
to show the test weight value. Use the
numeric keypad to enter the actual test weight, then press
to save the value and go to span calibration
(
WSPAN
).
5. With
WSPAN
displayed, press
to calibrate span. The indicator displays
*CAL*
while calibration is in
progress. When complete, the A/D count for the span calibration is displayed. Press
again to save the
span calibration value and go to the next prompt (
REZERO
).
6. The rezero function is used to remove a calibration offset when hooks or chains are used to hang the test
weights.
•
If no other apparatus was used to hang the test weights during calibration, remove the test weights and
press to return to the CALIBR menu.