18. The scale shows “EE-13
EE-13
EE-13
EE-13”. Press the zero key. The scale shows the current
initial zero range setting. Use the following table to select the “initial zero
range” setting:
Setting
Initial zero range
0
Off
1
On
A “0” setting implies: The initial zero setting mechanism (IZSM) will work
over the entire range of the scale capacity and the over-capacity point is 103%
above the zero obtained at calibration.
A “1” setting implies: The IZSM must be within +/- 10% of the zero
obtained at calibration and the over-capacity point is 103% above the IZSM.
19. The scale shows “EE-14
EE-14
EE-14
EE-14”. Press the zero key. The scale shows the current
graduation selection. Use the following table to select a graduation value.
Settings
Count by in lb
Count by in kg
0
100
50
1
50
20
2
20
10
3
10
5
4
5
2
5
2
1
6
1
0.5
7
0.5
0.2
8
0.2
0.1
9
0.1
0.05
10
0.05
0.02
11
0.02
0.01
12
0.01
0.01
NOTE: The stated accuracy specifications are based on the graduation setting in
the table below. If the graduation setting is set other than the value in the table
below the accuracy specification remains with the graduation size listed below.
If your capacity is:
Set your graduation to:
25 lb / 12.5 kg
.01 lb / .01 kg (12)
50 lb / 25 kg
.02 lb / .01 kg (11)
100 lb / 50 kg
.05 lb / .02 kg (10)
250 lb / 125 kg
.1 lb / 0.05 kg (9)
300 lb / 150 kg
.1 lb / 0.05 kg (9)
500 lb / 250 kg
0.2 lb / 0.1 kg (8)
1000 lb / 500 kg
0.5 lb / 0.2 kg (7)