Check for calibration blocking switch – Rice Lake Intercomp CS 750 - Users Manual User Manual
Page 15

CS750, Users
Rev R, May 2009
Page 15 of 27
9. The scale shows “EE-04
EE-04
EE-04
EE-04”. Press the zero key. Enter the auto off time in
minutes (0 to 255). The auto off time is how long the scale will remain ON
without any activity (a key being pressed or a change in weight). An entry of
0 turns the auto off feature OFF.
10. The scale shows “EE-05
EE-05
EE-05
EE-05”. Press the zero key. Enter what unit of measure
the scale should turn ON in; pounds or kilograms (0 to 1).
Setting
Units to turn ON in
0
pounds (lb)
1
kilograms (kg)
11. The scale shows”EE-06
EE-06
EE-06
EE-06”. Press the zero key. This function not used by
CS750.
12. The scale shows “EE-07
EE-07
EE-07
EE-07”. Press the zero key. Enter the sleep mode time in
minutes (0 to 255). The sleep mode time is how long the scale's display will
remain ON without any activity (a key being pressed or a change in weight).
An entry of 0 turns the sleep mode feature OFF. The sleep mode is designed
to conserve battery life on scales with a LED display.
13. The scale shows “EE-08
EE-08
EE-08
EE-08”. Press the zero key. Enter set point 1.
SET POINT 1
This is an optional feature that allows for an external set point. When the
specified weight (set point) is reached, a logic level high will be on the set
point connection. Set points are used with some other device (e.g. alarm,
relay) in conjunction to the CS750.
To activate a set point press the Set Point 1 (or Set Point 2) key. Enter
the weight you want the set point to activate.
Note: If the scale is over capacity (“OE
OE
OE
OE” will be displayed) the set points
will always become active.
SET POINT 2
Operates the same as Set Point 1.
14. The scale shows “EE-09
EE-09
EE-09
EE-09”. Press the zero key. Enter set point 2.
Check for calibration blocking switch
•
At this point the scale saves any changes that have been made.
•
A check is than made to see whether or not the calibration blocking switch is
enabled. If enabled, (CAL), the calibration procedure will continue. If disabled,
(RUN), the scale returns to normal weighing.