Rice Lake Counterpart Configurable Counting Indicator - User Manual Rev. D - Version 2.4 User Manual
Page 81

Serial Commands
75
SC.PRI.DSPDIV#n
Primary units display divisions
1D, 2D, 5D
SC.PRI.UNITS#n
Primary units
lb, kg, g, oz, OFF
SC.ACCUM#n
Accumulator enable
ON, OFF
SC.WZERO#n
Zero calibration
SC.WVAL#n
Test weight value
test_weight_value
SC.WSPAN#n
Span calibration
SC.WLIN.F1–
SC.WLIN.F5#n
Actual raw count value for linearization points 1–5
0–16777215
SC.WLIN.V1
SC.WLIN.V5#n
Test weight value for linearization points 1–5
0.000001–9999999
SC.WLIN.C1–
SC.WLIN.C5#n
Calibrate linearization points 1–5
SC.LC.CD#n
Deadload coefficient
SC.LC.CW#n
Span coefficient
SC.LC.CZ#n
Temporary zero
SC.REZERO#n
Rezero
SC.SEC#n
Secondary units
lb, kg, g, oz, OFF
SC.TER#n
Tertiary units
lb, kg, g, oz, OFF
SC.N#n
Report scale number being verified in Weigh Mode
Command
Description
Values
EDP.BAUD#p
Port baud rate
1200, 2400, 4800, 9600, 19200, 28800, 38400, 57600, 115200
EDP.BITS#p
Port data bits/parity
8NONE, 7EVEN, 7ODD
EDP.STOPBITS#p
Port stop bits
1, 2
EDP.PRNMSG#p
Print message
ON, OFF
EDP.ECHO#p
Port echo
ON, OFF
EDP.EOLDLY#p
Port end-of-line delay
0–255 (0.1-second intervals)
EDP.HEADERS#p
Port header
ON, OFF
EDP.INPUT#p
Port input
PORT 1 - CMD, SCANNER, SCALE, IND SC
PORT 2 - CMD, SCANNER
PORT 3 - CMD, SCANNER, KEYBOARD
EDP.PORT
Port reporting command. It responds to the port you are connected to.
EDP.RESPONSE#p
Port response
ON, OFF
EDP.SOURCE#p
Port source
1, 2, 3
EDP.STREAM#p
Port stream
OFF, LFT, INDUST
EDP.TERMIN#p
Port termination
CR/LF, CR
Table 6-5. Serial Port Serial Commands
Command
Description
Values
Table 6-4. Scales Serial Commands