Installation Manual
9.1.5
Normal Mode Commands
The normal mode print commands (see Table 9-11) transmit data to the serial port on demand in either setup or
normal mode.
Command
Description
Values
ALG.ALIAS#s
Analog output alias
name
ALG.SOURCE#s
Analog output source
PROG, SCALEn
ALG.MODE#s
Mode
GROSS, NET
ALG.OFFSET#s
Zero offset
0%, 20%
ALG.ERRACT#s
Error action
FULLSC, HOLD, ZEROSC
ALG.MIN#s
Minimum value tracked
0–9999999
ALG.MAX#s
Maximum value tracked
0–9999999
ALG.ZERO#s
Zero calibration
0–65535
ALG.SPAN#s
Span calibration
0–65535
For commands ending with “#s”, s is the slot number.
Table 9-9. ALGOUT Serial Commands (Valid Only If Analog Output Card Is Installed)
Command
Description
Values
XP#s
Extract probe temperature
—
XPP#s
Extract probe primary temperature
XPS#s
Extract probe secondary temperature
XPT#s
Extract probe tertiary temperature
XI#s
Extract 0–20 mA value
XV#s
Extract 0–10 V value
For commands ending with “#s”, s is the slot number.
Table 9-10. Analog Input Serial Commands (Valid Only If Analog Input Card Is Installed)
Command
Description
Values
CONSNUM
Set consecutive number
0–9 999 999
UID
Set unit ID
nnnnnnn
SD
Set date
MMDDYY, DDMMYY, YYMMDD, or YYDDMM. Enter six-digit
date using the year-month-day order specified for the
DATEFMT parameter, using only the last two digits of the year.
ST
Set time
hhmm (enter using 24-hour format)
SX#n
Start serial port streaming
OK or ??
The port streaming parameter (EDP.STREAM#p) for the
streaming port must be set to LFT or INDUST before using
these commands.
An EX command sent while in setup mode does not take
effect until the indicator is returned to normal mode.
EX#n
Stop serial port streaming
RS
Reset system
Soft reset. Used to reset the indicator without resetting the
configuration to the factory defaults.
Table 9-11. Normal Mode Serial Commands