Cp371 introduction – Kontron CP371 User Manual
Page 27
CP371
Introduction
ID 23762, Rev. 02
© 2002 PEP Modular Computers GmbH
Page 1 - 11
Note:
PRODUCTION ACCURACY 1: calibration by PEP; single samples
PRODUCTION ACCURACY 2: best case calibration (not supported by PEP); mean average
of samples (sample count > 100)
Table 1-5: CP371 Accuracy
INPUT
TYPE
and
RANGE
SIGNAL ACCURACY
at 23°C
± % of
FULL SCALE
ACCUMULATING
DEVIATION
% per K
(S and E1 BOARDS
ACCUMULATING
DEVIATION
% per K
(E2 BOARDS
BASIC AC
CURACY
0 V –
5 V
< 0.0488
< 0.0006
< 0.0006
0 V – 10 V
< 0.0244
< 0.0004
< 0.0004
- 5 V – + 5 V
< 0.0244
< 0.0004
< 0.0004
- 10 V – + 10 V
< 0.0244
< 0.0004
< 0.0004
0 mA – 20 mA
< 0.0488
< 0.0006
< 0.0006
BA
SIC A
CCURACY
(statistically evaluaated)
0 V –
5 V
< 0.0244
< 0.0004
< 0.0004
0 V – 10 V
< 0.0244
< 0.0004
< 0.0002
- 5 V – + 5 V
< 0.0244
< 0.0004
< 0.0002
- 10 V – + 10 V
< 0.0122
< 0.0002
< 0.0002
0 mA – 20 mA
< 0.0244
< 0.0004
< 0.0004
PRODUCTION
ACCURA
CY 1
VOLTAGE
< 0.15
< 0.001
< 0.001
CURRENT
< 0.15
< 0.001
< 0.001
PRODUCTION
ACCURAC
Y 2
VOLTAGE
< 0.25
< 0.001
< 0.001
CURRENT
< 0.25
< 0.001
< 0.001