46

|

Operating Protocols

HI-6113

L

ASER

D

ATA

I

NTERFACE

C

OMMANDS

HI-6113

Command

Description

Probe Response

i

Laser data interface

identification string

:i6113

sr=10-character software

revision

sn=8-character serial

number

n

Read laser current

:nx.xxx.

o

Laser OFF command

:o

The laser and all LEDs except

the green Power LED will turn

off

r

Laser ON command

:r

The blue Laser LED will

illuminate immediately, then the

yellow Receive LED will

illuminate a few seconds after,

indicating the probe is ready for

operation

The Laser ON command, r, should be the first command sent.

HI-6113

Command

Description

Probe Response

tc

Read temperature in

Centigrade

:txxxx.

tf

Read temperature in

Fahrenheit

:txxxx.