22

|

HI-6113 Indicators

The HI-6113 Laser Data Interface contains the laser for powering the

Field Probe, and also handles the communication between the probe and the

host computer.

The first two LEDs on the front of the device indicate the status of the laser:

LASER HOT (red)

-

Constant On

This is a warning that the laser is running too hot

and the system should be checked. Dirty fiber ends, poor inline

connections, or damaged fiber cables are among the causes of a

LASER HOT warning. The system should be shut down as soon

as possible, and the cause of the additional heat investigated.

Continuing to run in a LASER HOT state will reduce the life of the

HI-6113.

-

Flash

This is an indication that the laser current is approaching

0.8 mA. During the initialization phase of the startup process the

HI-6113 provides the probe with more light and then rolls back the

power to a normal level. In these instances the user may see

LASER HOT flash on; this is normal operation.

LASER OK (blue)

This is the normal status when probe

communication is established and the HI-6113 is providing laser light

to the Field Probe for power.

RECEIVE (amber) and TRANSMIT (white)

These are illuminated

during normal use. RECEIVE indicates the communication between

the Field Probe and the HI-6113, and TRANSMIT indicates the

communication between the host computer and the HI-6113.

POWER (Green)

This indicates that the HI-6113 is using power from

the DC power supply. This should be lit as soon as the HI-6113 is