ProbeView Laser Software

M

OVING

A

VERAGE

ProbeView Laser may be set to provide a moving average of up to 50 samples. A

setting of 1 indicates no averaging.

T

EXT UPDATE RATE

Allows the user to set the text update rate in the Probe-Status-Box to a readable

rate. With an HI-61XX Series Field Probe, ProbeView Laser can acquire more

than 70 samples per second. If every sample were displayed in text format, fast

sample rates would become unreadable.

A

NNOTATING

L

OGGED

D

ATA

ProbeView Laser will prompt the user for information about the logged data that

is saved. These prompts are configurable. To configure the prompts:

Select the Options window.

Click the Misc tab.

Select Edit Note Prompts. The prompt, prompt-identifier, and

last-user-entry are stored as a single line in the file.

The prompt is the text preceding the first comma.

The text preceding the first comma is the prompt-identifier, which is written to

the saved data file with the user response to the prompt.

The last-user-entry is the last response the user gave to the prompt. It is also

saved in the prompt file.