Moving average, Text update rate, Annotating logged data – ETS-Lindgren HI-6113 Laser Data Interface and Probe Measurement System User Manual
Page 36

36
|
ProbeView Laser Software
M
OVING
A
VERAGE
ProbeView Laser may be set to provide a moving average of up to 50 samples. A
setting of 1 indicates no averaging.
T
EXT UPDATE RATE
Allows the user to set the text update rate in the Probe-Status-Box to a readable
rate. With an HI-61XX Series Field Probe, ProbeView Laser can acquire more
than 70 samples per second. If every sample were displayed in text format, fast
sample rates would become unreadable.
A
NNOTATING
L
OGGED
D
ATA
ProbeView Laser will prompt the user for information about the logged data that
is saved. These prompts are configurable. To configure the prompts:
Select the Options window.
Click the Misc tab.
Select Edit Note Prompts. The prompt, prompt-identifier, and
last-user-entry are stored as a single line in the file.
The prompt is the text preceding the first comma.
The text preceding the first comma is the prompt-identifier, which is written to
the saved data file with the user response to the prompt.
The last-user-entry is the last response the user gave to the prompt. It is also
saved in the prompt file.