beautypg.com

Taylor-Wharton XL-55HP (with dual regulators) User Manual

Page 13

background image

GAS SERVICE CHANGE KITS

Kit

Valve

Connection

Part No.

Gas Service

Name

Designation

GL50-8C35

Oxygen

LIQUID

CGA 440

VENT

CGA 440

USE

CGA 540

GL50-8C30

Nitrogen

LIQUID

CGA 295

VENT

CGA 295

USE

CGA 580

GL50-8C31

Argon

LIQUID

CGA 295

VENT

CGA 295

USE

CGA 580

HP50-8C30

Carbon Dioxide

LIQUID

CGA 320

VENT

CGA 295

USE

CGA 320

HP50-8C35

Nitrous Oxide

LIQUID

CGA 326

VENT

CGA 295

USE

CGA 326

REGULATOR MAINTENANCE
A dual stage, spring loaded regulator is employed for the pressure building/economizer

circuit. This regulator can be adjusted on the container, replaced or checked and adjusted

off the container in a readily fabricated bench adjustment fixture.

Regulator Adjustment – On Container
1. Fill the container with the appropriate liquid product.
2. Open the Pressure Building Valve and allow the container pressure to stabilize for

about an hour. Note the point where the pressure stabilizes.

3. Adjust the screw on the top of the regulator to raise or lower the pressure to the

desired point. When decreasing the setting, the pressure building valve must be closed

and the container vented to a lower pressure. Then repeat step 2 in order to observe

the change.

REGULATOR ADJUSTMENT RANGES

Part No.

Normal Setting

Range

Delta

8816-1060

400 psig

300 to 600 psig

28 bar

20.7 to 41 bar

2758 kPa

2068 to 4137 kPa

6999-9018

300 psig

200 to 350 psig

20 psig

20.7 bar

13.8 to 24.1 bar

1.4 bar

2068 kPa

1379 to 2413 kPa

138 kPa

6999-9015

125 psig

75 to 175 psig

8.6 bar

5 to 12 bar

862 kPa

517 to 1207 kPa

CAUTION

When changing gas

service, install the

proper fittings – DO NOT

use adapters. The

following procedures

address the physical

changes to the container

only. For detailed

procedures on the

decontamination of the

container itself, refer to

CGA pamphlet C-10

“Changes of Service for

Cylinders Including

Procedures for

Inspection and

Contaminant Removal.”

CAUTION

Carbon Dioxide and

Nitrous Oxide may

contain contaminants

such as hydrocarbons,

that are not easily

removed from cylinders,

and associated with

components by

conventional oxygen

service cleaning

procedures. Once a

cylinder is placed into

CO

2

or N

2

O gas service, it

should never be

converted to another gas

service. See CGA

pamphlet C-10 for proper

procedures.

NOTE:

One clockwise turn of

the adjustment will raise

the setpoint by

approximately 30 psig (2

bar/207 kPa). See the

chart below to

determine the range of

adjustment for the

regulator you are

servicing. Do not attempt

to set the regulator to a

pressure outside of its

design range.

13

This manual is related to the following products: