Installation, Operation, Getting unit into service – Taylor-Wharton KRYOS Controller User Manual

Page 9: Electrical, Power supply connection, Initial fill, Control components

background image

7

KRYOS

If possible, the original packing material should be retained for reshipment. If not
available, consult Taylor-Wharton for shipping and packing instructions. It is the
responsibility of the customer to assure that the goods are adequately packaged for
return to the factory. All refrigerators returned to Taylor-Wharton must be clean and
sterile before return. See page 13 for cleaning instructions.

INSTALLATION

Getting Unit into Service

Your Cryostorage System should come with complete instructions for how you should
remove the unit from the crate and put it into service. Read both this manual and
your Cryostorage System’s manual before beginning any installation. Make sure to
follow any required procedures and safety guidelines when you are hooking up your
Liquid Nitrogen source.

Electrical

The liquid level controllers used with these refrigerators operate from 24/16.5 VAC.
However, the external transformer does have a 110/220 VAC primary. Disconnect
the electrical power cord from the outlet before attempting any service.

Power Supply Connection

Connect the power supply to your LABS or K-Series Cryostorage System and then
plug the power supply into a surge protected 110/220 VAC outlet.

OPERATION

These instructions are for operators experienced with cryogenic equipment. Before
operating the system, become familiar with the safety precautions in this manual
and in reference publications. Make certain all applicable provisions set forth in the
Installation Section have been followed before placing a system in operation. Study
this manual thoroughly. Know the location and function of all system components.

Initial Fill

The Cryostorage System using the KRYOS controller, comes preset from the factory.
The liquid nitrogen supply pressure at the inlet to the refrigerator should be in
the range of 10 psig (0.7 bar/69 kPa) to 20 psig (1.4 bar/138 kPa) for optimum
performance. Higher operating pressures will increase transfer losses and create
excessive turbulence of the liquid in the refrigerator which can generate false signals
to the liquid level controller causing the refrigerator to under fill. In “liquid phase”
storage applications, excessive turbulence can cause splashing which could result in
personal injury.

Control Components

Interface Panel

The KRYOS Interface panel, which the user will interact with, contains the vacuum
fluorescent display as well as the number keypad, power button, help button and the
soft-key control buttons. (See Figure 2.0 and 2.1 on following page.)

WARNING:

Electrical shock can

kill. Do not attempt

any service on these

units without first

disconnecting the

electrical power cord.

CAUTION:

If using a power

source other than

North American

110/220 VAC,

contact Taylor-

Wharton Customer

Service to make

sure you are using

the appropriate

converter. Use of an

unapproved power

source converter may

cause permanent

damage to the unit.

NOTE:
For KRYOS Field
Installation instructions
refer to Page 35.

WARNING:

Maintain adequate

ventilation to

prevent asphyxiation

hazard (see Safety

Precautions).

WARNING:

If the fill fails to

stop for any reason,

quickly close the

liquid supply valve

to prevent overfilling

until the cause of

the problem can be

determined.