Factory Settings

To restore factory settings at any time KRYOS can be re-initialized. Please note that a re-initialization clears all
logs and passwords and resets settings to those listed in the table below. To reset factory settings in the control,
go to the menu choice ABOUT THIS CONTROL. (

MENU, 6, 4) and while looking at this screen, type in 1973.

The control will lead you through the process required to reset the factory settings.

Temperature:

Thermocouple #1 Enabled

Thermocouple #2 Disabled

Hi Temp #1 Alarm -100º C

Hi Temp #2 Alarm Disabled

Level Control:

For LABS 20/38/40/80

OFFSET 0 inches

START FILL Sensor 3 inches

STOP FILL Sensor 5 inches

Sensor Assembly Freeze-Guard

K-Series and LABS 94

OFFSET 0 inches

START FILL Sensor 3 in. (76.2 mm)

STOP FILL Sensor 6 in. (152.4)

Sensor Assembly Freeze-Guard

System:

RS-232 Enabled

Handshaking CTS/RTS

Baud Rate 19200

Date & Time USA Central Time

Brightness 100%

Lid Switch Enabled

Auto Defog Timer 20 seconds

Quick-chill 15 seconds

Temperature Control:

Temperature Control Disabled

Temperature ºC

Temperature Range ±ºC

Alarms:

LN

2

Supply Alarm 30 minutes

Sensor Error Enabled

Remote Alarm 30 minutes

Lid Open Too Long 10 minutes

Thermocouple Open Alarm Enabled

Logging:

Sensor Error Log Enabled

Low LN

2

Enabled

Supply Log Enabled

Remote Alarm Log Enabled

Open Thermocouple Enabled

Hi Temp #1 Log Enabled

Hi Temp #2 Log Disabled

Fill Logging Enabled

Lid Open/Closed Log Enabled

User Access Log Disabled

Thermocouple #1 Rate Log Enabled

Thermocouple #2 Rate Log Disabled

Freeze-Guard:

Valve Open Duration Disabled

Valve De-icing Enabled

REPLACEMENT PARTS

A complete list of Replacement Parts and Accessories for the LABS and K-Series
Cryostorage Systems is available from the following Taylor Wharton Customer
Care Centers:

Europe

Americas

Asia/Australia

xHFEjNkIIgKFKgKEIK

xFIEgHHGgLJLN˜ 0TT#0�#&#ULNNgLMLgFJIK

xJEjNkFJNHNgFIGG

36

KRYOS

Factory Settings

To restore factory settings at any time KRYOS can be re-initialized. Please note that a re-initialization clears all
logs and passwords and resets settings to those listed in the table below. To reset factory settings in the control,
go to the menu choice ABOUT THIS CONTROL. (

MENU, 6, 4) and while looking at this screen, type in 1973.

The control will lead you through the process required to reset the factory settings.

Temperature:

Thermocouple #1 Enabled

Thermocouple #2 Disabled

Hi Temp #1 Alarm -100º C

Hi Temp #2 Alarm Disabled

Level Control:

For LABS 20/38/40/80

OFFSET 0 inches

START FILL Sensor 3 inches

STOP FILL Sensor 5 inches

Sensor Assembly Freeze-Guard

K-Series and LABS 94

OFFSET 0 inches

START FILL Sensor 3 in. (76.2 mm)

STOP FILL Sensor 6 in. (152.4)

Sensor Assembly Freeze-Guard

System:

RS-232 Enabled

Handshaking CTS/RTS

Baud Rate 19200

Date & Time USA Central Time

Brightness 100%

Lid Switch Enabled

Auto Defog Timer 20 seconds

Quick-chill 15 seconds

Temperature Control:

Temperature Control Disabled

Temperature ºC

Temperature Range ±ºC

Alarms:

LN

2

Supply Alarm 30 minutes

Sensor Error Enabled

Remote Alarm 30 minutes

Lid Open Too Long 10 minutes

Thermocouple Open Alarm Enabled

Logging:

Sensor Error Log Enabled

Low LN

2

Enabled

Supply Log Enabled

Remote Alarm Log Enabled

Open Thermocouple Enabled

Hi Temp #1 Log Enabled

Hi Temp #2 Log Disabled

Fill Logging Enabled

Lid Open/Closed Log Enabled

User Access Log Disabled

Thermocouple #1 Rate Log Enabled

Thermocouple #2 Rate Log Disabled

Freeze-Guard:

Valve Open Duration Disabled

Valve De-icing Enabled

REPLACEMENT PARTS

A complete list of Replacement Parts and Accessories for the LABS and K-Series
Cryostorage Systems is available from the following Taylor Wharton Customer
Care Centers:

Europe

Americas

Asia/Australia

xHFEjNkIIgKFKgKEIK

xFIEgHHGgLJLN˜ 0TT#0�#&#ULNNgLMLgFJIK

xJEjNkFJNHNgFIGG

36

KRYOS

Factory Settings

To restore factory settings at any time KRYOS can be re-initialized. Please note that a re-initialization clears all
logs and passwords and resets settings to those listed in the table below. To reset factory settings in the control,
go to the menu choice ABOUT THIS CONTROL. (

MENU, 6, 4) and while looking at this screen, type in 1973.

The control will lead you through the process required to reset the factory settings.

Temperature:

Thermocouple #1 Enabled

Thermocouple #2 Disabled

Hi Temp #1 Alarm -100º C

Hi Temp #2 Alarm Disabled

Level Control:

For LABS 20/38/40/80

OFFSET 0 inches

START FILL Sensor 3 inches

STOP FILL Sensor 5 inches

Sensor Assembly Freeze-Guard

K-Series and LABS 94

OFFSET 0 inches

START FILL Sensor 3 in. (76.2 mm)

STOP FILL Sensor 6 in. (152.4)

Sensor Assembly Freeze-Guard

System:

RS-232 Enabled

Handshaking CTS/RTS

Baud Rate 19200

Date & Time USA Central Time

Brightness 100%

Lid Switch Enabled

Auto Defog Timer 20 seconds

Quick-chill 15 seconds

Temperature Control:

Temperature Control Disabled

Temperature ºC

Temperature Range ±ºC

Alarms:

LN

2

Supply Alarm 30 minutes

Sensor Error Enabled

Remote Alarm 30 minutes

Lid Open Too Long 10 minutes

Thermocouple Open Alarm Enabled

Logging:

Sensor Error Log Enabled

Low LN

2

Enabled

Supply Log Enabled

Remote Alarm Log Enabled

Open Thermocouple Enabled

Hi Temp #1 Log Enabled

Hi Temp #2 Log Disabled

Fill Logging Enabled

Lid Open/Closed Log Enabled

User Access Log Disabled

Thermocouple #1 Rate Log Enabled

Thermocouple #2 Rate Log Disabled

Freeze-Guard:

Valve Open Duration Disabled

Valve De-icing Enabled

REPLACEMENT PARTS

A complete list of Replacement Parts and Accessories for the LABS and K-Series
Cryostorage Systems is available from the following Taylor Wharton Customer
Care Centers:

Europe

Americas

Asia/Australia

xHFEjNkIIgKFKgKEIK

xFIEgHHGgLJLN˜ 0TT#0�#&#ULNNgLMLgFJIK

xJEjNkFJNHNgFIGG