System logs, Temperature logs, Erase logs – Taylor-Wharton KRYOS Controller User Manual
Page 21: Security, User options

19
KRYOS
System Logs
FILL LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 1). This menu choice turns on/off the logging of tank filling
operations. The choices are ENABLE or DISABLE. Records data in the system log.
LID ACTION LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 2). This menu choice turns on/off the logging of lid openings and
closings. The choices are ENABLE or DISABLE. Records data in the system log.
USER ACCESS LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 3). This menu choice turns on/off the logging of user access codes,
which are requested when the lid is opened. Records data in the system log.
Temperature Logs
Temperature Logging Rates for thermocouple #1 (T/C #1 LOGGING) and
thermocouple #2 (T/C #2 LOGGING) are accessible through the menu system of
the control (
MENU, 4, 4, 1 or MENU, 4, 4, 2). This menu choice adjusts the rate
at which temperatures are logged for the two thermocouples. The possible choices
are Disabled, 15, 30 minutes, 1, 2, 4, 6, 12 or 24 hours. Records data in the
temperature logs.
Erase Logs
ERASE LOGS is accessible through the menu system of the control (
MENU, 4, 5).
This menu choice erases any of the four logs found in the control. Please note
that once a log has been erased, it is gone forever.
ERASE SYSTEM LOGS (
MENU, 4, 5, 1).
ERASE ERROR LOGS (
MENU, 4, 5, 2).
ERASE TEMPERATURE LOG #1 (
MENU, 4, 5, 3).
ERASE TEMPERATURE LOG #2 (
MENU, 4, 5 ,4).
Security
KRYOS security features restrict access to certain key features such as power and the
menu system.
The POWER-ON PASSWORD can be set through the menu system of the control
(
MENU, 5, 1). The power-on password requires entry of a 4-digit password before
turning on or turning off the control. Follow the directions on the display to set a new
password or disable a password. Codes 0000 & 9999 are invalid choices. Code
9999 is used to access a procedure to clear passwords should they be forgotten.
The MENU PASSWORD can be set through the menu system of the control
(
MENU, 5, 2). The menu password requires entry of a 4-digit password before
allowing access to the menu system to change any control settings. Follow the
directions on the display to set a new password or to disable a password. Codes
0000 & 9999 are invalid choices. A code of 9999 is used to access a procedure to
clear passwords should they be forgotten.
User Options
The USER OPTIONS menu choice covers all other control settings not already
covered. These include serial port settings, date and time settings, Lid Settings,
Defog, Quick-Chill, Temperature Control, Valve Freeze-Guard, Display brightness
and Control Information.
19
KRYOS
System Logs
FILL LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 1). This menu choice turns on/off the logging of tank filling
operations. The choices are ENABLE or DISABLE. Records data in the system log.
LID ACTION LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 2). This menu choice turns on/off the logging of lid openings and
closings. The choices are ENABLE or DISABLE. Records data in the system log.
USER ACCESS LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 3). This menu choice turns on/off the logging of user access codes,
which are requested when the lid is opened. Records data in the system log.
Temperature Logs
Temperature Logging Rates for thermocouple #1 (T/C #1 LOGGING) and
thermocouple #2 (T/C #2 LOGGING) are accessible through the menu system of
the control (
MENU, 4, 4, 1 or MENU, 4, 4, 2). This menu choice adjusts the rate
at which temperatures are logged for the two thermocouples. The possible choices
are Disabled, 15, 30 minutes, 1, 2, 4, 6, 12 or 24 hours. Records data in the
temperature logs.
Erase Logs
ERASE LOGS is accessible through the menu system of the control (
MENU, 4, 5).
This menu choice erases any of the four logs found in the control. Please note
that once a log has been erased, it is gone forever.
ERASE SYSTEM LOGS (
MENU, 4, 5, 1).
ERASE ERROR LOGS (
MENU, 4, 5, 2).
ERASE TEMPERATURE LOG #1 (
MENU, 4, 5, 3).
ERASE TEMPERATURE LOG #2 (
MENU, 4, 5 ,4).
Security
KRYOS security features restrict access to certain key features such as power and the
menu system.
The POWER-ON PASSWORD can be set through the menu system of the control
(
MENU, 5, 1). The power-on password requires entry of a 4-digit password before
turning on or turning off the control. Follow the directions on the display to set a new
password or disable a password. Codes 0000 & 9999 are invalid choices. Code
9999 is used to access a procedure to clear passwords should they be forgotten.
The MENU PASSWORD can be set through the menu system of the control
(
MENU, 5, 2). The menu password requires entry of a 4-digit password before
allowing access to the menu system to change any control settings. Follow the
directions on the display to set a new password or to disable a password. Codes
0000 & 9999 are invalid choices. A code of 9999 is used to access a procedure to
clear passwords should they be forgotten.
User Options
The USER OPTIONS menu choice covers all other control settings not already
covered. These include serial port settings, date and time settings, Lid Settings,
Defog, Quick-Chill, Temperature Control, Valve Freeze-Guard, Display brightness
and Control Information.
19
KRYOS
System Logs
FILL LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 1). This menu choice turns on/off the logging of tank filling
operations. The choices are ENABLE or DISABLE. Records data in the system log.
LID ACTION LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 2). This menu choice turns on/off the logging of lid openings and
closings. The choices are ENABLE or DISABLE. Records data in the system log.
USER ACCESS LOGGING is accessible through the menu system of the control
(
MENU, 4, 3, 3). This menu choice turns on/off the logging of user access codes,
which are requested when the lid is opened. Records data in the system log.
Temperature Logs
Temperature Logging Rates for thermocouple #1 (T/C #1 LOGGING) and
thermocouple #2 (T/C #2 LOGGING) are accessible through the menu system of
the control (
MENU, 4, 4, 1 or MENU, 4, 4, 2). This menu choice adjusts the rate
at which temperatures are logged for the two thermocouples. The possible choices
are Disabled, 15, 30 minutes, 1, 2, 4, 6, 12 or 24 hours. Records data in the
temperature logs.
Erase Logs
ERASE LOGS is accessible through the menu system of the control (
MENU, 4, 5).
This menu choice erases any of the four logs found in the control. Please note
that once a log has been erased, it is gone forever.
ERASE SYSTEM LOGS (
MENU, 4, 5, 1).
ERASE ERROR LOGS (
MENU, 4, 5, 2).
ERASE TEMPERATURE LOG #1 (
MENU, 4, 5, 3).
ERASE TEMPERATURE LOG #2 (
MENU, 4, 5 ,4).
Security
KRYOS security features restrict access to certain key features such as power and the
menu system.
The POWER-ON PASSWORD can be set through the menu system of the control
(
MENU, 5, 1). The power-on password requires entry of a 4-digit password before
turning on or turning off the control. Follow the directions on the display to set a new
password or disable a password. Codes 0000 & 9999 are invalid choices. Code
9999 is used to access a procedure to clear passwords should they be forgotten.
The MENU PASSWORD can be set through the menu system of the control
(
MENU, 5, 2). The menu password requires entry of a 4-digit password before
allowing access to the menu system to change any control settings. Follow the
directions on the display to set a new password or to disable a password. Codes
0000 & 9999 are invalid choices. A code of 9999 is used to access a procedure to
clear passwords should they be forgotten.
User Options
The USER OPTIONS menu choice covers all other control settings not already
covered. These include serial port settings, date and time settings, Lid Settings,
Defog, Quick-Chill, Temperature Control, Valve Freeze-Guard, Display brightness
and Control Information.