beautypg.com

Baumer CombiTemp User Manual

Page 2

background image

Page 2

Design and specifi cations subject to change without notice

Installation Manual 5850-008

www.baumerprocess.com

Environmental conditions

Media temperature, std.

-50...400°C

Surface sensor

-40...150°C

Ambient temperature

-40...160°C

(or max. temperature range for display/transmitter)

Humidity

< 100% RH, condensing

(or max. humidity for display/transmitter)

Protection class

DIN housing IP 65

ø80 mm housing IP 65 + IP 66

Vibrations

GL, test 2

(sensor tubes < 200 mm only)

Disposal of product and packing

According to national laws or by returning to Baumer

Sensor tube and connection

Material

Acid-proof, stainless steel

(AISI

316L/W.1.4404)

Media pressure

Max. 16 bar

Time constant t

0.5

See table below

Mechanical tolerances

ISO 2768-m

Sensor element

Sensor type

Pt100, Class A or B
Pt1000, Class B

Accuracy DIN/EN/IEC

60751

1/1 DIN B:

±(0.3 + 0.005 x t) °C

1/3 DIN B:

+1/3 x (0.3 + 0.005 x t) °C

1/6 DIN B:

+1/6 x (0.3 + 0.005 x t) °C

1/1 DIN A:

+(0.15 + 0.002 x t) °C

Technical Data, General

Safety Instructions

Ex-Application - Confi guration

Valid for FlexTop 2201 / 2211 / 2221 and FlexView:

None of the two types of FlexProgrammer confi guring unit must be
connected to the FlexView within the hazardous area.

Confi guring procedure:
a) Disconnect mains from the 4...20 mA loop circuit.
b) Disconnect the product from the circuitry within

the hazardous area.

c) Bring the product to the safe area.
d) Connect the FlexProgrammer and perform the
confi

guring

session.

e) Re-install the product in the hazardous area.
f) Connect the power supply to the circuit.

Valid for FlexTop 2221 only:
Confi guration of the FlexTop2221 can be made within the hazardous
area by means of a handheld HART confi gurator, providing the
precautions and guidelines described in the product´s manual are
observed.

Valid for FlexTop 2231 only:
Confi guration of the FlexTop2231 can be made within the hazardous
area by means of a Profi bus PA confi gurator, providing the precautions
and guidelines described in the product´s manual are observed.

This instrument is built and tested ac cord ing to the current EU-direc-
tives and packed in technically safe con di tion. In order to main tain this
condition and to ensure safe op er a tion, the user must follow the hints
and warnings given in this instruction.

During the installation the valid national rules have to be observed.
Ignoring the warnings may lead to severe per son al injury or substantial
damage to property.

The product must be op er at ed by trained staff. Correct and safe
operation of this equipment is de pend ent on proper transport, storage,
installation and op er a tion.

All electrical wiring must conform to local stand ards. In order to prevent
stray elec tri cal radiation, we rec om mend twisted and shielded input
cables, as also to keep power sup ply cables separated from the input
cables. The con nec tion must be made ac cord ing to the connecting
di a grams.

Before switching on the power supply take care that other equip ment
is not af fect ed. Ensure that the supply voltage and the conditions in the
environment comply with the spec i fi ca tion of the de vice.

Before switching off the supply voltage check the possible effects on
other equipment and the processing system.

WARNING

For electrical installation and com mis sion ing of explosion protected
devices, the data given in the con form i ty cer tifi cate as also the local
reg u la tions for installation of elec tri cal apparatus within explosion pro-
tect ed areas must be considered. The intrinsically safe versions can be
mounted in the explosion hazarded area according to its spec i fi ca tion
only con nect ed to a certifi ed intrinsically safe supply loop with the cor-
re spond ing electrical values.

WARNING

This product contains no replaceable parts.
In case of malfunction the product must be shipped to Baumer for
repair.

WARNING

Products with painted and/or plastic surfaces (e.g. display) imply a risk
of electrostatic charging.
To prevent electrostatic hazard - do only clean with a moist cloth.
After mounting the device - do check that the housing has a ground
potential.