Running liquid controls – ITC Avoximeter 4000 Whole Blood CO-Oximeter User Manual
Page 48
ITC AVOXimeter 4000 Operator’s Manual
●
RNA CO-Oximeter controls, available from RNA Medical, Devens, MA., phone 978 772-
●
Multi-4™ CO-Oximeter controls, available from Instrumentation Laboratory, Inc.,
1.
Verify that the instrument is ready to run a test and that the “READY” - “Insert
Cuvette”
screen is displayed:
Running Liquid Controls
ITC recommends either of the following Liquid Quality Controls (LQC):
9070 or toll-free 800-533-6162.
Lexington, MA., phone toll-free 800-955-9525.
2.
Fill a test cuvette with a liquid control (see page 32).
3.
Insert the test cuvette into the test hamber (see page 33).
c
4.
If optional entry of User IDs is enabled (see page 26), the screen for entering a
User ID is displayed:
Enter the User ID, then press
. When the confirmation screen is displayed,
press
followed by
.
Note:
To bypass entry of a User ID, either press
, enter a zero and then
press
, or simply press
when the screen for entering a User
ID is displayed.
5.
The screen for specifying the sample type is displayed:
followed by
6.
Press
. The screen for specifying QC type is displayed:
followed by
7.
Press
. The screen for specifying the control level is
displayed:
40