Instrument specifications, Reportable range, Accuracy – ITC Avoximeter 4000 Whole Blood CO-Oximeter User Manual

Page 18: Precision

background image

ITC AVOXimeter 4000 Operator’s Manual

Instrument Specifications

Size

20.3 cm (8.0 in) x 25.4 cm (10.0 in) x 9.5 cm (3.8 in)

Weight

1.8 kg (4 lbs)

Operating Temperature

Room temperature (15°C to 30°C, 59°F to 86°F)

Battery Type

Nickel Cadmium (NiCad)

Operating Time On Battery

Approximately 8 hours (constant run) or 10 complete test cycles
per charge. Tests may also be run while the ITC AVOXimeter
4000

is plugged into the AC/DC power module.

Anticipated Battery Life

Approximately 500 charge / discharge cycles

Power Supply/Chargers

Input:

100 / 240 VAC, 50 / 60 Hz

Output: 12 VDC, 800 mA

Serial Data Port

RS232C

Sample Type

Whole blood

Sample Volume

50 µL

Analysis Time

7 to 10 seconds per sample

Analysis Wavelengths

5

Reportable Range

tHb

4 to 25 g/dL

%O

2

Hb

0 to 100%

%COHb

0 to 75%

%MetHb

0 to 85%

[O

2

]

0 to 35 mL O

2

/dL

Accuracy

tHb (>10 g/dL)

±0.45 g/dL

tHb (<10 g/dL)

±0.35 g/dL

%O

2

Hb

±1.6%

%COHb

±2%

%MetHb

±1.5%

Precision

tHb

0.3 g/dL

%O

2

Hb

0.8 % O

2

Hb

%COHb

1 % COHb

%MetHb

0.7 % MetHb

10