Running a test, Running a test on a patient sample – ITC Avoximeter 4000 Whole Blood CO-Oximeter User Manual
Page 41
ITC AVOXimeter 4000 Operator’s Manual
Running a Test
Running a Test on a Patient Sample
1.
Verify that the instrument is ready to run a test and that the “READY” - “Insert
Cuvette”
screen is displayed:
2.
Holding the cuvette by the finger grip on the black cap, insert the cuvette (with the
syringe still attached) into the test chamber.
Important:
Always keep the syringe attached when inserting the cuvette into the
accurate results.
test chamber. Removing the syringe may cause in
CAUTION: Never inject sample directly into the test chamber.
CAUTION: Carefully review the additional precautions on page 8.
3.
If optional entry of User IDs is enabled (see page 26), the screen for entering a
User ID is displayed:
Enter the User ID, then press
. When the confirmation screen is displayed,
press
followed by
.
r ID, either press
Note:
To bypass entry of a Use
, enter a zero and then
press
, or simply press
when the screen for entering a User
ID is displayed.
33