Specifying a different value for hüfner’s number – ITC Avoximeter 4000 Whole Blood CO-Oximeter User Manual
Page 36
![background image](/manuals/602983/36/background.png)
ITC AVOXimeter 4000 Operator’s Manual
Specifying a Different Value for Hüfner’s Number
f oxygen that can be carried by one gram of hemoglobin) is used
t of a sample (see page 1). Although Hüfner’s number is generally
assumed to be 1.39 mL/gm, facilities may use a different value and may wish to set the ITC
AVOXimeter 4000
to match other instruments in the facility.
1.
Display the “Calibration” menu (a submenu of
Hüfner’s number (the volume o
to calculate the oxygen conten
, see page 7).
2.
Press
followed by
to display the screen for accep g
tin or changing the
value used for Hüfner’s number:
followed by
3.
Press
to display the screen for entering a different value:
Note:
A default value of 1.39 is entered at the factory. Values ranging from
1.30 to 1.39 can be entered.
4.
Press the number key (e.g.,
) corresponding to the new value:
5.
Check that the new Hüfner’s numb r is correctly displayed, then press
e
followed by
. The “Calibration” menu is again displayed.
6.
Press
to return to the previous menu, if desired.
28