Press

(RUN TEST) You should see this screen:

Press

(MARSHALL TEST) This will automatically start the platen to

rise at a rate of 2.0 in/min (50.8mm/min). You should see this screen:

The display above will give constant real-time readings of the stability and
flow measurements during the test cycle. You can end the test at any time
by pressing

. This will stop all data collection and reverse the platen

to its down position.

As the breaking head makes contact with the load cell, the HM-3000
will automatically zero the stability and flow readings and begin to takes
readings every 0.1 seconds. Once the test has reached peak stability (i.e.
peak load) and fallen back 2%, data will cease being collected and the platen
will begin to reverse and move back to the down position. As the platen
moves down, the screen will display the peak values.

You should see this screen:

* REVIEW LAST TEST
* RUN TEST

* MORE SETUP

* MARSHALL TEST
* CBR/LBR TEST
* USER DEFINED TEST

MARSHALL TEST # of readings
STAB

“#”

FLOW

“#”

* END TEST

STAB PEAK

“#”

FLOW PEAK

“#”

REVIEW DATA
* RUN MARSHALL TEST