0 troubleshooting – Bio-Rad Econo Gradient Pump Kits User Manual
Page 39

8.0 TROUBLESHOOTING
Problem
Possible Cause
Solution
Pump displays Power outage
If you wish to continue a run in
“Power Failure!”
progress, press the Continue
soft key. The LCD will display
the program at the time of the
power failure. Pressing the
Stop soft key will display the
message “Program Aborted”.
You may now operate the
pump
as normal.
Pump will
No power
Verify that power cord is
not run
(no LEDs lit)
plugged in and the power
switch is turned on.
Fuses
Check that proper fuses are
installed and intact.
Insure that the calibration
indicator light is not flashing.
For pump to run, this light must
either be off or steadily lit.
Mode
Econo Gradient Pump must be
in either Manual or Program
mode to start.
Liquid remains Platen not
Adjust the platen by turning
stationary
adjusted
it counter clockwise until it is
in tubing
completely loosened, then
rotate three to five full turns
clockwise. See section 4.3,
Platen Adjustment.
Clogged tubing,
Check tubing, fittings, and
valves, or fittings
valves for obstructions.
Damaged tubing
Check pump tubing. Replace
damaged tubing.
Troubleshooting
35
4006161B.qxd 8/14/2003 8:09 AM Page 35
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- ChromLab™ Software (178 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)