Section 11
Troubleshooting Guide
Problem Cause
Solution
Optics validation
Problem with the
Repeat the procedure.
procedure shows
optical component of
If value is still out of
value outside of
the array reader
range, contact Bio-Rad
acceptable range
technical support.
Reporter validation
Problem with the
Repeat the procedure.
procedure shows
optical component of
If value is still out of
value outside of
the array reader
range, contact Bio-Rad
acceptable range
technical support.
Classify validation
Problem with the
Repeat the procedure.
procedure shows
calibration or
If values are still out of
value outside of
optical component
range, contact Bio-Rad
acceptable range
of the array reader
technical support.
Fluidics validation
Problem with fluidics
Repeat procedure. If
procedure shows
lines, valves or
value is still out of range,
value outside
sample needle of
contact Bio-Rad
of acceptable range
array reader
technical support.
Section 12
Ordering Information
Catalog #
Description
171-203001
Bio-Plex Validation Kit 4.0, includes optics validation,
reporter validation, classify validation, and fluidics validation bead
sets for approximately 50 calibration routines
171-203060
Bio-Plex Calibration Kit, includes Cal1 and Cal2 calibration
beads for approximately 50 daily calibration routines
171-203032
Bio-Plex MCV Plate III, for use with Bio-Plex Manager 4.0 or
4.1 (standard or security software) and validation kit 4.0
171-203033
Bio-Plex MCV Plate IV, for use with Bio-Plex Manager 5.0
(standard or security software) and later versions and validation
kit 4.0
21
4110185B.qxp 9/25/2007 11:55 AM Page 21