Section 11
Troubleshooting Guide

Problem Cause

Solution

Optics validation

Problem with the

Repeat the procedure.

procedure shows

optical component of

If value is still out of

value outside of

the array reader

range, contact Bio-Rad

acceptable range

technical support.

Reporter validation

Problem with the

Repeat the procedure.

procedure shows

optical component of

If value is still out of

value outside of

the array reader

range, contact Bio-Rad

acceptable range

technical support.

Classify validation

Problem with the

Repeat the procedure.

procedure shows

calibration or

If values are still out of

value outside of

optical component

range, contact Bio-Rad

acceptable range

of the array reader

technical support.

Fluidics validation

Problem with fluidics

Repeat procedure. If

procedure shows

lines, valves or

value is still out of range,

value outside

sample needle of

contact Bio-Rad

of acceptable range

array reader

technical support.

Section 12
Ordering Information

Catalog #

Description

171-203001

Bio-Plex Validation Kit 4.0, includes optics validation,
reporter validation, classify validation, and fluidics validation bead
sets for approximately 50 calibration routines

171-203060

Bio-Plex Calibration Kit, includes Cal1 and Cal2 calibration
beads for approximately 50 daily calibration routines

171-203032

Bio-Plex MCV Plate III, for use with Bio-Plex Manager 4.0 or
4.1 (standard or security software) and validation kit 4.0

171-203033

Bio-Plex MCV Plate IV, for use with Bio-Plex Manager 5.0
(standard or security software) and later versions and validation
kit 4.0