Suspension array system validation report – Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual
Page 22
![background image](/manuals/600307/22/background.png)
your product insert for values specific to your product control number..
Parameter
Specification
Measured Value
Pass/Fail
DD Median
4774 - 6593
6133
Pass
CL1 Median
3383 - 4135
3615
Pass
CL1 CV%
2.00 - 7.00
3.52%
Pass
CL2 Median
3520 - 4302
3883
Pass
CL2 CV%
3.00 - 8.00
6.93%
Pass
RP1 Median
3509 - 4290
3878
Pass
RP1 CV%
4.00 - 10.00
8.89%
Pass
Parameter
Specification
Measured Value
Pass/Fail
% Carryover
< or = 4.0%
1.3%
Pass
A. Calculated Values
B. Raw Values
Parameter
Specification
Measured Value
Pass/Fail
Bead
Median FI
Dynamic Range
4.33 - 4.45
4.40
Pass
Blank
2
Linearity
> 0.995
1.000
Pass
1
8
Slope 33.94
-
39.42
37.57
Pass
2
54
Accuracy
> 90.00%
96.94%
Pass
3
656
Threshold
< 6 MFI
2
Pass
4
1576
5
4453
6
25059
A. Classify Efficiency
Classify Bead
Specification
Measured Value
Pass/Fail
Bead A (1)
> 80.0%
98.1%
Pass
Bead B (4)
> 80.0%
93.5%
Pass
Bead C (40)
> 80.0%
90.5%
Pass
Bead D (54)
> 80.0%
95.3%
Pass
Bead E (100)
> 80.0%
93.0%
Pass
B. DD Efficiency
Parameter
Specification
Measured Value
Pass/Fail
DD Efficiency
> or = 75.0%
96.7%
Pass
Comments:
Reviewed by:
Date:
III. Reporter Validation
IV. Classify Validation
II. Fluidics Validation
Test Performed by: JPocekay
Date/Time: 04-Aug-04 11:55 AM
Expiration Date: 15-Oct-06
Reader Serial #: LX10000266002
Access Level: Unrestricted
Validation Kit Control #: 999999 (Low)
RP1 Target Value: 3885
RP1 PMT Voltage: 571.67
Result: Passed
Result: Passed
Result: Passed
Result: Passed
I.Optics Validation
Suspension Array System Validation Report
20
4110185B.qxp 9/25/2007 11:55 AM Page 20