Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 13

background image

7.

Place 5 drops of each bead into the respective wells on the MCV plate (see
Figure 2 below or Figure 1B). Store beads at 4°C as soon as possible after
use. Protect the beads from light.

Fig. 2. MCV plate IV.

8.

Select the E

Ejje

ec

ctt button in the dialog box to eject the plate holder.

9.

Place the MCV plate in the microplate platform.

10. Select O

OK

K to begin all validation procedures.

11. When the procedure has completed, the results will be displayed in a dialog

box, as shown in Figure 3:

Fig. 3. Validation results.