Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 11

background image

3.

If the control numbers on the BizCard CD already exist in the database, you
will receive the message that the control numbers are already installed and
the program will exit. Click on the OK button and remove the BizCard CD.

4.

If the control numbers on the BizCard CD are not found in the database, you
will see a dialog asking you to press OK to install the control numbers with
their specifications. Click on the OK button.

5.

A dialog box will appear informing you that control numbers installation is
complete. You may now proceed with validation in Bio-Plex Manager using
the current control number.

Section 10
Procedure for Performing Validation

Introduction

This section provides instructions for use of the Bio-Plex

®

validation kit 4.0 using

either Bio-Plex Manager™ 4.0/4.1 software with MCV plate III or Bio-Plex
Manager 5.0 and later with MCV plate IV. Either software version provides a fully
automated validation routine that sequentially performs all the validation tests
without further user intervention. To perform all validations in a single step, follow
instructions in Section 10.1. If you wish to perform an individual validation routine,
consult Sections 10.2–10.5.

N

No

otte

e:: IIn

n tth

he

e ffo

ollllo

ow

wiin

ng

g iin

ns

sttrru

uc

cttiio

on

ns

s a

allll rre

effe

erre

en

nc

ce

es

s tto

o tth

he

e M

MC

CV

V p

plla

atte

e w

wiitth

ho

ou

utt

s

sp

pe

ec

ciiffy

yiin

ng

g IIIIII o

orr IIV

V iin

nd

diic

ca

atte

e tth

he

e p

prro

oc

ce

ed

du

urre

e iis

s u

us

se

ed

d ffo

orr e

eiitth

he

err c

co

on

nffiig

gu

urra

attiio

on

n.. A

Allll

p

piic

cttu

urre

es

s a

arre

e u

up

pd

da

atte

ed

d tto

o rre

efflle

ec

ctt tth

he

e n

ne

ew

we

es

stt M

MC

CV

V p

plla

atte

e,, b

bu

utt d

do

oe

es

s n

no

ott iim

mp

plly

y tth

ha

att

tth

he

erre

e iis

s a

an

ny

y d

diiffffe

erre

en

nc

ce

e iin

n p

pe

errffo

orrm

ma

an

nc

ce

e..

10.1 One-Step Procedure for all Validation Parameters

Procedure

1.

Turn on the Bio-Plex array reader, microplate platform, and computer as
specified in the Bio-Plex hardware and Bio-Plex Manager user manuals.

2.

Perform start-up procedure as directed.

3.

Calibrate Bio-Plex array reader using Cal1 and Cal2 beads found in the
Bio-Plex calibration kit according to the Bio-Plex Manager software manual.

N

No

otte

e:: B

Be

e s

su

urre

e tto

o c

ca

alliib

brra

atte

e iim

mm

me

ed

diia

atte

elly

y b

be

effo

orre

e v

va

alliid

da

attiio

on

n.. IIff y

yo

ou

u a

arre

e u

us

siin

ng

g B

Biio

o--P

Plle

ex

x

M

Ma

an

na

ag

ge

err 4

4..0

0//4

4..1

1,, u

us

se

e e

eiitth

he

err tth

he

e h

hiig

gh

h o

orr llo

ow

w R

RP

P1

1 tta

arrg

ge

ett ffo

orr C

Ca

all2

2 c

ca

alliib

brra

attiio

on

n

d

de

ep

pe

en

nd

diin

ng

g o

on

n tth

he

e tty

yp

pe

e o

off a

as

ss

sa

ay

y b

be

eiin

ng

g p

pe

errffo

orrm

me

ed

d.. IIff y

yo

ou

u a

arre

e u

us

siin

ng

g B

Biio

o--P

Plle

ex

x

M

Ma

an

na

ag

ge

err 5

5..0

0,, tth

he

e c

ca

alliib

brra

attiio

on

n iis

s d

do

on

ne

e o

on

n llo

ow

w R

RP

P1

1 tta

arrg

ge

ett.. S

Se

elle

ec

cttiio

on

n o

off h

hiig

gh

h R

RP

P1

1

tta

arrg

ge

ett iis

s d

do

on

ne

e iin

n tth

he

e v

va

alliid

da

attiio

on

n s

sc

crre

ee

en

n..

4.

Remove all validation bead sets from 4°C storage and vortex each bottle for
30 sec. This is very important for proper validation.

9

4110185B.qxp 9/25/2007 11:55 AM Page 9