10. When the procedure is completed, values will be displayed in a dialog box as
shown below (Figure 11). The classify efficiency and DD efficiency results
may be accessed in this view.
Fig. 11. Classify validation results.
11. The results will also be logged into a validation log in Bio-Plex Manager
software. To access this log, select V
Viie
ew
w from the main menu then V
Va
alliid
da
attiio
on
n
L
Lo
og
g. Each type of validation is stored in a separate log for the purpose of
tracking data over time. See the Bio-Plex Manager software user guide for
your version of software for more information on using the validation log.
12.
If any values do not meet specifications, repeat the procedure. If values are
again not within specifications, contact Bio-Rad technical service for
assistance.
10.6 Generating a Validation Report
Procedure
The results from each validation procedure are sent to a validation log in Bio-Plex
Manager software. This log may be used to create individual reports as well as
track multiple validation results over time. Each type of validation is logged into a
separate view: optics validation, fluidics validation, reporter validation and classify
validation. You may maneuver through each of the views using either the main
menu items or the toolbar icons. The specifications for each control number of
the validation kit are also shown in a separate window below the results. If All
Validation was selected, an entry matching the specific date and time will
appear in each of the validation logs. All of the validation results for a specific
date and time will be included in a created report. The entire log may be printed
by selecting P
Prriin
ntt then R
Re
es
su
ulltts
s from the main menu.
A general procedure for creating a report from the validation log is shown below.
For more detailed instructions on the use of the validation log, consult the
Bio-Plex Manager user guide.