Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 19

background image

12. Repeat procedure if values are not within the specification range. If values

are again not within acceptable specification ranges, contact Bio-Rad
technical support for assistance.

10.5 Validation of Classify Efficiency

Procedure

1.

If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system.

N

No

otte

e:: B

Be

e s

su

urre

e tto

o c

ca

alliib

brra

atte

e iim

mm

me

ed

diia

atte

elly

y b

be

effo

orre

e v

va

alliid

da

attiio

on

n.. IIff y

yo

ou

u a

arre

e u

us

siin

ng

g B

Biio

o--P

Plle

ex

x

M

Ma

an

na

ag

ge

err 4

4..0

0//4

4..1

1,, u

us

se

e e

eiitth

he

err tth

he

e h

hiig

gh

h o

orr llo

ow

w R

RP

P1

1 tta

arrg

ge

ett ffo

orr C

Ca

all2

2 c

ca

alliib

brra

attiio

on

n

d

de

ep

pe

en

nd

diin

ng

g o

on

n tth

he

e tty

yp

pe

e o

off a

as

ss

sa

ay

y b

be

eiin

ng

g p

pe

errffo

orrm

me

ed

d.. IIff y

yo

ou

u a

arre

e u

us

siin

ng

g B

Biio

o--P

Plle

ex

x

M

Ma

an

na

ag

ge

err 5

5..0

0,, tth

he

e c

ca

alliib

brra

attiio

on

n iis

s d

do

on

ne

e o

on

n llo

ow

w R

RP

P1

1 tta

arrg

ge

ett.. S

Se

elle

ec

cttiio

on

n o

off h

hiig

gh

h R

RP

P1

1

tta

arrg

ge

ett iis

s d

do

on

ne

e iin

n tth

he

e v

va

alliid

da

attiio

on

n s

sc

crre

ee

en

n..

2.

Remove the classify validation bead set from 4°C storage and vortex each
bottle for 30 sec.

3.

Place 5 drops of each classify bead into the corresponding classify well
labeled as A, B, C, D, and E in the MCV plate.

4.

Store stock vials at 4°C as soon as possible after use. Protect beads from
light.

5.

Select IIn

ns

sttrru

um

me

en

ntt from the main menu. Select V

Va

alliid

da

attiio

on

n from the pull-down

menu.

6.

Enter the User name and select control number. Select C

Clla

as

ss

siiffy

y V

Va

alliid

da

attiio

on

n.

Select OK. The following dialog box will appear (Figure 10):

Fig. 10. Classify validation dialog.

7.

Select the Eject button to eject the plate holder.

8.

Place the MCV plate on the microplate platform.

9.

Select O

OK

K to start the classify validation procedure.

17

4110185B.qxp 9/25/2007 11:55 AM Page 17