Section 9
Procedure for Bio-Plex Manager 3.0 and
MCV plate II

Introduction

This section provides instructions for use of the Bio-Plex Validation Kit 3.0 with
Bio-Plex Manager version 3.0 and MCV plate II. Bio-Plex Manager now provides
a fully automated validation routine that sequentially performs all the validation
tests without further user intervention. To perform all validations in a single step,
follow instructions in Section 9.1. If you wish to perform an individual validation
routine, these instructions are provided in Sections 9.2–9.5.

9.1 One-Step Procedure for all Validation Parameters

Procedure

1.

Turn on the Bio-Plex array reader, microplate platform, and computer as
specified in the Bio-Plex hardware and Bio-Plex Manager user manuals.

2.

Perform start-up procedure as directed.

3.

Calibrate Bio-Plex array reader using CAL 1 and CAL2 beads found in the
Bio-Plex Calibration Kit according to the Bio-Plex Manager software manual.

N

No

otte

e:: B

Be

e s

su

urre

e tto

o c

ca

alliib

brra

atte

e iim

mm

me

ed

diia

atte

elly

y b

be

effo

orre

e v

va

alliid

da

attiio

on

n.. U

Us

se

e tth

he

e H

Hiig

gh

h R

RP

P1

1

T

Ta

arrg

ge

ett ffo

orr C

CA

AL

L2

2 c

ca

alliib

brra

attiio

on

n..

4.

Remove all validation bead sets from 4°C storage and vortex each and every
bottle for 30 sec. This is very important for proper validation.

Bio-Plex Manager 3.0 and MCV Plate II Instructions

9

4110007c.qxd 02/25/2003 10:11 AM Page 9