Bio-Rad Gel Doc™ XR+ System User Manual
Page 38

38
The following window will open with a notification that the system will be acquiring reference files,
depending on the imager:
a. Gel Doc XR
Note: No lights should be turned ON. Please allow approximately 30 seconds for the acquisition
of the reference file.
b. ChemiDoc XRS
This process takes a fraction of a second.
After the reference file acquisition is over the system will be ready to operate.
1. Turn ON the lights using the Epi-white touch pad button.
2. Place the focusing target on the transilluminator platen.
3. Click on Live/Focus.
4. An image of the target will be seen on the imaging screen.
5. Using the lens control buttons on the membrane touch pad or the control panel that appears in the
Quantity One acquisition window, adjust the iris, focus, and zoom.
Note: For additional details on Reference File Acquisition, refer to the Quantity One software
manual.