4 Silver Stain Plus

Reagent Preparation

Development Accelerator Solution

1. Prepare once for use with the entire contents of the

kit.

2. Place 950 ml of deionized distilled water in a

1 L cylinder containing a PTFE-coated stir bar.

3. Begin stirring and slowly add the entire

50 g contents of the development accelerator reagent.
Stir until dissolved.

4. Add water to make 1 L. Pour the solution into the

provided empty 1 L bottle labeled development
accelerator solution.

5. Store at 4°C. Use at room temperature.

If the entire solution will not be used within 3 months,
prepare the development accelerator solution according to
your use rate.

To prepare half the solution, add 25 g development
accelerator reagent to 475 ml deionized distilled water,
add water to make 500 ml.