10.3 Measuring the luminance

candela / m

2

(cd/m

2

) or footLambert (fL)


- Set the measuring head to reflected light measuring

or

. The luminance function is set and

shown in the display.
- Focus on the subject to be measured via the
viewfinder
- Take a measurement using the measuring button

M

The selected display unit and measured luminance
are displayed.






10.4 Measuring time-integral values

(Ixs, fcs, cds/m

2

, fLs)

- Set measuring head to reflected light or incident light
measuring. The corresponding display unit is set and
shown in the display.
- Set with the right menu button the function
flash light .
- Pre-select the desired measuring time using
the setting wheel; this can be found on the right hand
side of the display panel. Activated half time stop
increments are also displayed.
- Start flash measuring using the measuring button

M

Cord/noncord (chapter 6.1, page 23).
- The measured value, which is calculated to
1 second, is shown in the pre-selected display unit.
By preselecting the measuring time, the ambient
light part is correspondingly taken into account.