28 Gossen Foto- und Lichtmesstechnik GmbH

8.1.1 Measuring correction values


Correction values can also be measured directly.
An evenly illuminated surface and constant light level
are required.
Use the STARLITE 2 in the reflected light mode at
1 ° or 5°.
In the function

, a reference measurement can be

taken by pressing the measuring button

M

.

Designation rF - - in the digital display.
Then, hold the filter in front of the viewfinder and press
the measuring button

AVR

.

The light reducing effect will be indicated automatically in
the display as EV stop and extension factor.




8.1.2 Deleting correction values

In the function

- correction values (chapter 8.1,

page 27) you have two possibilities of deleting pre-
programmed correction values:
- by manually resetting using the setting wheel to
EV 0 and extension factor 1.0 or
- by pressing the measuring button

M

(display rF - -)
- quit the correction value function using the

menu button ( or )

Correction value is deleted.
Symbol

has disappeared from the display.