From the menu bar, click on Diagnostics,
then Instrument Calibration.

Vision requests that the aperture and Tablet
Transmission Standard Tray be inserted, and
that “Use Tray” be selected from the
Configure, Tray menu.

These items are reviewed below. Be sure to
select “Tablet Transmission Standard” as the
tray type. Click “Use Tray” and “Select”.

If all items have been done, and the tablet transmission instrument
reference is in place, please click “OK”.

Wavelength Linearization is performed in two sections,
each of which takes about 45 seconds.

No user response is required during this test.